Beckman DxC 700 AU

https://www.myco-instrumentation.com/web/image/product.template/83473/image_1920?unique=1425089

User-friendly, reliable, high throughput analyzer for mid to high-volume clinical labs.

MYCO provides expert installation and repair services.

Ask us about our multi-year service contracts, reagent contracts, and selling your old instruments.

In Stock

0.00 0.0 USD 0.00

0.00

This combination does not exist.

Get your quote in 90 minutes.
Trade in your older analyzers to recover budget.
Bundle equipment, reagents, & consumables into one cost-saving order.
Enlist our local service tech to keep uptime high.
Lower your cost per test.

Product Code: B98654

Manufacturer: Beckman Coulter

WHY CHOOSE MYCO INSTRUMENTATION?
See Why in 2 mins.

Need replacement parts for your DxC 700 AU? We manufacture them. Click here to find them quickly.

Beckman DxC 700 AU lowers operational costs for busy labs


Hospitals and high-volume diagnostic labs are always looking for ways to get faster turn-around time while keeping operational costs low.  

The DxC 700 AU Offers:

Low Total Cost of Ownership
Few Service Calls
Low Cost Per Test
Simple Software

Most Lab Technicians have experiences with older Beckman analyzers, the DxC 700 AU gives you only the best features of its predecessors:  

 User-Friendly Software 

The DxC 700 AU software is nothing like the older AU series. This interface is truly intuitive and easy to use.


 High Throughput

800 photometric tests per hour, up to 1,200 with ISE. Ideal for mid to high-volume labs.

 Load Samples and Reagents Anytime

Cooled STAT compartment and continuous racking loader to give you faster turnaround time. 

 Affordable Replacement Parts

Fewer consumable parts to lower total cost of ownership.

Easy DxC 700 AU software for uninterrupted productivity


No more software challenges! 

Your technicians would also be glad to know that the famously challenging AU-series software is gone. 

Your staff will have a very low learning curve with the DxC 700 AU interface. 

Your department will increase in testing capacity and recover the DxC 700 AU investment faster.

"The new software is easy to learn and simple to operate with very little experience."

Dustin Steger, B.S. (Chemistry) • Marshfield Labs®, Wisconsin


Request a quote now

Monitor analysis, reagents levels, and generate reports in one centralized platform



Faster turn-around time with DxC 700 AU


Run urgent samples STAT and load reagents mid-operation

There is a cooled STAT compartment in the DxC 700 AU.

The 150- sample continuous rack loader lets you load samples any time.

You can also load reagents while the analyzer is running.


Request a quote now

"The ability to load reagents 'on-the-fly' while the instrument is running saves me time and minimizes workflow disruption."

Cassidy LeMay, B.S. (Chemistry) • Marshfield Labs, Wisconsin.

Fewer service calls for the DxC 700 AU


Higher uptime and longer average repair intervals than other analyzers in the same cost class. 

Your lab wastes less budget on external service, since most troubleshooting on the DxC 700 AU can be handled in house. 

More uptimes for mid to high-volume labs with DxC 700 A

Automatic Clot Detection

Identifies clots that could clog probes or tubing, preventing sample processing errors.

Probe Crash Protection

Detects probe crashes before they cause significant damage or inaccuracies.

The cost of running your DxC 700 AU is lower than the most common analyzer in its class, the Siemens Visa 500. 
Read the Independent Study Here: Evaluation of the Beckman Coulter DxC 700 AU chemistry analyzer.


See your quote in 90 minutes


More accuracy with the DxC 700 AU sample quality analysis


The Lipemia, Hemolysis, and Icterus Indices identify compromised samples quickly, preventing inaccurate results due to interference.

Lipemia Index

Measures the level of fat in the blood, which can interfere with test results.

Hemolysis Index

Detects the rupture of red blood cells, which can falsely elevate certain analytes.

Icterus Index


Identifies elevated bilirubin levels, which may skew results in some assays.

The built-in sample quality analysis minimizes errors and increases result reliability. That is more process efficiency for your lab.


DxC 700 AU in high humidity


Validated performance between 20% - 80% RH 

If your environment is humid, this is the analyzer for you.

The DxC 700 AU routinely outperforms other chemistry analyzes in humid environments, with significantly fewer errors.


Beckman DxC 700 AU tip!

Give your lab a productivity boost by pairing your system with a high-quality DI water system. 

The DxC 700 AU performs better than most in the class when you keep it hydrated.


Request a quote now


Reduce DxC 700 AU consumable expenses


Compared to other systems in its class, the DxC 700 AU requires fewer consumables. The concentrated reagents, durable ion-selective electrodes (ISEs), and reusable cuvettes help you lower the cost per test.

If your consumable budget is based on OEM prices, here’s the best part: you save a chuck of that cost instantly by ordering replacement parts from MYCO.
How?

We manufacture these parts in-house. Our parts are designed to meet or surpass OEM specifications, ensuring seamless compatibility with your chemistry analyzer. 

You save 50% on reagent probes and sample probes, and Save 25-30% on:

  Cuvettes

  Syringes

  Optical lamps

  Seals and Tubing

  OL Filters

  Wash Buffer

  Maintenance Kits

You can order your parts from us directly, and we’ll ship them out the same day.


Request a quote now


Fund your DxC 700 AU upgrade with your old instruments


Sell us your old instruments and the cost of your Beckman DxC 700 AU becomes even more manageable. We make equipment sales easy for you. 

Guaranteed Value

We give you a competitive value on your equipment based on set parameters 

Safe Transaction

You get a clear payment timeframe and no hidden clauses

Hassle-Free Process

We assist with the decontamination plus we arrange
and pay for all shipping changes. Your process is hassle free.

We assist with the decontamination plus we arrange and pay for all shipping changes. Your process is hassle free.


DxC 700 AU service & repairs


Nationwide, onsite service is available (even though your DxC 700 AU will rarely need it).
We can give you:

  • Customized on-site annual service contracts
  • Professional installation
  • Equipment familiarization following your installation

Time and material service visits, if your DxC 700 AU is not under our service contract.


Contact our service team

MYCO offers quick and responsive services for Beckman Coulter analyzers.

Beckman DxC 700 AU tip!


Keep downtimes to a minimum with one (or a few) service kit on hand. You won’t regret it, especially for the same quality and compatible kit that costs 25% less than OEM. 

Are you a biomedical equipment service provider?

Tell us how many of the DxC 700 AU's are in your care, and we can give you volume pricing on parts to keep on hand for your upcoming repair and maintenance needs.


Order service kits here 


We want you to be happy and less stressed at work


MYCO has served clinical labs across the United States and around the world for more than 40 years. We know how hard your days can be. 
So let us help you make them easier.

Your Beckman DxC 700 AU is ready to ship when you need it. 

We also offer comprehensive installation, training, and repair services on every chemistry analyzer we carry.


Request a quote today


Instruments Specifications

  • Newly designed intuitive graphical user interface, including simplified reagent inventory, calibration and quality control procedures

  • Online maintenance documentation

  • User-customized menu and free-text notepad

  • Colour alerts to highlight system operating conditions

  • Reliability for greater uptime with quick and easy replacement of common parts

  • No special tools required

  • No more than three steps and no longer than 60 seconds for replacement of parts, such as sample and reagent probes, mixers, syringes, ISE electrodes and optical lamp

  • Extensive online help with instructional videos
  • Available integration with pre-analytical automation through direct track sampling from Power Express

  • Available connection to Command Central with REMISOL Advance, allowing access to multiple instruments and automation consoles from a single workstation, optimizing laboratory management and improving decision making

  • PROService remote diagnostic capabilities ensure 24/7 continuous monitoring of vital system functions

  • Cooled STAT compartment, providing one-button STAT interrupt, advanced auto QC and calibration capabilities

  • Ability to load reagents while the analyzer is in measure mode

  • Priority sample repeat/reflex

  • Economical ISEs with long onboard stability; easy to maintain (only individual electrode replacement required)

  • 150-sample continuous rack loader

More Specifications

Analytical system


Fully automated, random-access clinical chemistry system with STAT capability


Analytical principles

Spectrophotometry and potentiometry


Assay types

Endpoint, rate, fixed point and indirect ISE


Analytical methods

Colorimetry, turbidimetry, latex agglutination and indirect ISE


Test menu applications

I20


Programmable tests

117 photometrics (113 + LIH + HbA1c (Thb, HbA1c + HbA1c%)) and ISE


Onboard parameters

60 photometric tests and 3 ISEs (Na, K, Cl)


Throughput

800 photometric tests per hour, up to 1,200 with ISE

ISE sample throughput: 200 per hour

ISE maximum tests per hour: 600 if ISE only


Sample types

Serum, plasma, urine, whole blood (HbA1c) and other fluids


Sampler capacity

Rack sampler: 10 samples per rack (barcodes on primary tubes and on racks), capacity for 150 samples

Refrigerated STAT carousel: 22 samples can be run simultaneously (cal, QC and routine samples)


Sample tubes

Primary and secondary tubes: diameter between 11.5 mm and 16 mm; height between 55 mm and 102 mm; nested micro cups


Sample volume

μL-25 μL in 0.1 μL increments


Sample quality analysis

Lipemia, hemolysis and icterus indices

Clot detection and probe crash protection

Sample bar-code formats


NW7, EAN13, CODE 39, CODE 128, ISBT-128, 2 of 5 standard, 2 of 5 interleaved mixed readable (max. four types at the same time, except if using ISBT-128)


Reagent supply

60 positions for R1, 48 positions for R2 (refrigerated 4°C-12°C); bottle sizes: 15 mL, 30 mL, 60 mL


Reagent volume

R1: 10 μL-250 μL; R2: 10 μL-250 μL (1 μL increments)


Total reaction volume

120-350 μL


Reaction cuvette

Permanent glass cuvettes


Reaction time

Up to 8 minutes, 33 seconds


Reaction temperature

37°C ±0.3°C


Reaction method

Dry bath


Photometric range

0 OD-3.0 OD


Wavelength

13 different wavelengths between 340 nm and 800 nm


Calibration

Auto-calibration, advanced calibration and cooled calibrator positions

Calibrator value assignment by 2D barcode; 200 calibrators can be programed

History of graphical calibration data stored


Quality control

Westgard Rules, Twin Plot and Levey-Jennings graphs, auto QC, cooled QC positions

100 controls can be programed, 10 levels per test

Reflex testing


User-defined

Automated sample pre-dilution

Repeat run with increased or decreased sample volume or sample pre-dilution (3, 5, 10, 15, 20, 25, 50, 75, up to 100 times)


Online

Uni- and bi-directional host query communications


Operating system

Windows 7


Data storage

Up to 100,000 patient samples; reaction monitor: 400,000 tests, 300 index


INSTALLATION REQUIREMENTS

Dimensions (W x H x D)


Analyzer:

49 in x 51 in x 35 in (1250 mm x 1300 mm x 890 mm) 1,015 lbs (460 kg)


Sampler:

29 in x 37 in x 41 in (730 mm x 940 mm x 1040 mm)

331 lbs (150 kg)


Power supply

200V, 208V, 220V, 230V, 240V, 50 Hz, 60 Hz, 3.8 kVA


Water supply information

Mean water consumption: 28 L per hour

Water type: deionized CAP Type II, bacteria free

Continuous flow supply

Resistivity: less than 2.0 μS/cm filtered with a 0.5 μm filter


Temperature and humidity

18°C to 32°C, 20% to 80% RH (no condensation)


Drain requirements

Built-in waste pump


Drain required: maximum height from floor < 59 in (< 1.5 m)

Technical Specs


• Photometric Throughput with ISE (tests/hour) 


• Number of Onboard Assays with ISE


• Sample Volume (µL)


• Sample Load Capacity (Routine/STAT)


• Clot Detection and Auto-Clearing (Mechanism)


• Open Channel Capability 


• Software 



• 800/1,200 with ISE


• 63


• 1.0-25.0

• 150 (15 racks x 10 samples), 22 samples by carousel.


• Yes


• Yes


• Windows 10

EQUIPMENT REFURBISHED TO PERFORM 'LIKE NEW', IT IS NOT A 1:1 SUBSTITUTE FOR NEW AND DOES NOT FALL UNDER MANUFACTURER'S WARRANTY.